Rosemount 1151GP5E22S1B1 Pressure Sensor with1199PFW11A11. Condition is Used. Removed in good working condition from a ongoing operation
Performance Specifications:
Rosemount 1151
(Zero-based calibrated ranges, reference conditions, silicone oil fill, 316 SST isolating diaphragms for HART 4-20 mA protocol.)
Accuracy:
Output Model Accuracy Specification and Span
Output Code S Ranges 3 through 8 for DP and GP; ±0.075% of calibrated span between 1:1 to 10:1 of URL
Ranges 4 through 7 for HP URL
0.02 --- – 0.1 % of calibrated span between 10:1 and 50:1 of URL span
URL
Square Root Mode 0.2 + 0.05 x--- % of calibrated flow span for all spans
span
All other ranges and transmitters ±0.25% of calibrated span for all spans
Output Codes Ranges 3 through 5 for DP and GP ±0.2% of calibrated span for all spans E, G, L, and M
P8 Option: Ranges 3 through 8 for DP ±0.1% of calibrated span for > 10 inH2O and GP, all HP and all LT
All other ranges and transmitters ±0.25% of calibrated span for all spans
Stability
Output Code: S L and M
Temperature Effect: Output Code
S Line Pressure Effect Model
DP Range 4 and 5
DP Transmitters Ranges 6 - 0
Model
Ranges 3-8
All ranges
Model
DP/GP Ranges 4-8, HP Ranges 4-8
Zero Error
±0.25% of URL for 2,000 psi (13790 kPa), correctable through rezeroing at line pressure.
±0.5%, correctable through rezeroing at line pressure.
Specification
±0.1 of URL for 2 years
±0.25 of URL for 6 months
Specification
Zero Error = ±0.2% URL per 100 °F (56 °C)
Total Error = ±(0.2% URL + 0.18% of calibrated span) per 100 °F; double the effect for other ranges and transmitters
Span Error
Correctable to ±0.25% of input reading per 1,000 psi (6895 kPa)
Correctable to ±0.25% of input reading per 1,000 psi (6895 kPa)
E and G Ranges 3-6 ±0.2 of URL for 6 months All other ranges ±0.25 of URL for 6 months
E, G, L, and M Ranges 4-0 Zero Error = ±0.5% URL per 100 °F.
Total Error = ±(0.5% URL + 0.5% of calibrated span) per 100 °F; double the effect for Range 3.
DP Range 3 ±0.5%, correctable through rezeroing at line Correctable to ±0.5% of input reading per 1,000 psi (6895 kPa) pressure.
HP Transmitters ±2.0% of URL for 4,500 psi (31027 kPa), Correctable to ±0.25% of input reading per 1,000 psi (6895 kPa).
Rosemount™ 1199 Diaphragm Seal System
The Rosemount 1199 Diaphragm Seal System offers protection to transmitters, guarding against hot and cold, corrosive or erosive processes. This seal system features a variety of solutions, including industry-specific seals for difficult processes. This safety certified system does not require mounting hardware and can be used for many pressure measurement applications, providing reliable, remote measurements.