To get consistent results, you need smooth flow delivery and correct gradient composition. The Vanquish Binary Pump H has exceptional low pressure ripple, contributing to incredible retention time stability. Minimal pulsation ensures that the same flow and gradient are delivered in each run for consistent results every time.
Manufacture Specification
Flow Accuracy: 0.1%
GLP Features:
Predictive Performance functions for scheduling maintenance procedures based on the actual operating and usage conditions of the pump.
All system parameters logged in the Chromeleon Chromatography Data System Audit Trail Flow Precision: <0.05% RSD or <0.01 min SD, whichever is greater
Gradient Formation: High-pressure gradient proportioning
Pressure Range: 5 to 151 MPa, (50 to 1517 bar, 700 to 22,000 psi)
I/O Interfaces: 2 x 6 pin Mini-DIN connectors each having functionality: 1 input, 1 relay out, 1 bidirectional input/outpt
No. of Solvents: 2 out of 6
Proportioning Accuracy: 0.2% of full-scale (from 0.2 to 4.0 mL/min and 1 to 99%)
Proportioning Precision: <0.15% SD (from 0.2 to 4.0 mL/min and 1% to 99%)
Wetted Parts: MP35 N, DLC, titanium, zirconium oxide (ZrO2), sapphire, aluminium oxide (Al2O3), PEEK, PTFE, ECTFE, FEP, UHMW polyethylene, perfluoro-elastomer, amorphous fluoropolymer (AF)
Compressibility Compensation: Compensation fully automated, independent of mobile phase composition
PC Connection: USB 2.0 (3-port-HUB to connect further Vanquish modules)
Safety Features: Leak detection and safe leak handling, excess pressure monitoring
Temperature Range: 5C to 35C (20 to 80% RH (non-condensing))
Pressure Ripple: <0.4% or <0.2 MPa, whichever is greater (Typically <0.2% or <0.05 MPa, whichever is greater)
Voltage: 100-240 V AC, 50/60 Hz, max. 525 W/550 VA
Weight (English): 70.5 lb.
Weight (Metric): 32 kg
Description:
Mixer Volume: 25 L (default configuration)
Biocompatible Version: Yes, pH range 2-12, buffer and/or chloride concentration up to 1 mol/L
Max. Flow Rate: 5 mL/min
Min. Flow Rate: 0.001 mL/min
Type: Parallel Dual Piston with Independent Piston Drives and Variable Stroke Volume
Additional Prep Fees May Apply