Mettler Toledo Thornton Conductivity/Resistivity Instrument 200CR
**This unit is in box ready to be use. New in box unused in open box. Includes everything in pictures. See photos for further details.
The 200CRS is an analytical and process control instrument for measuring solution properties. It is user friendly, flexible and highly accurate. The 200CRS processes the signal from one sensor. A 1 line x 16 character Liquid Crystal Display (LCD) module conveys measuring data and setup information. The display is backlit for viewing in all lighting conditions. The menu structure allows the operator to modify all operational parameters by using keys on the front panel. A menu-lockout feature, with password protection, is available to prevent the unauthorized use of the meter. The 200CRS is configured with two relays for process control. The 200CRS instrument is equipped with a communication interface that can be configured as either an RS422 or an RS232. This interface provides real-time data output and complete instrument configuration capabilities for central monitoring via Personal Computer or Programmable Logic Controller.
FEATURES
Display: 1 line x 16 character backlit LCD.
Measurements: resistivity, conductivity, C, F, total dissolved solids.
Measurement Channels: 1.
Signal Inputs per channel: 2 (total of 2 signals for measurement).
Measurement Cycle Time: 1 second (4 measurements processed per second).
Configuration: all setup information is stored in non-volatile memory.
Setpoints (alarms): 2 independent alarms programmable as high or low limits.
Relays: 2 with programmable delay time and hysteresis.
Outputs: 2 analog outputs (4-20mA), rescalable to 0-20mA.
Communications: RS232/RS422 interface, bidirectional.
Security: keypad lockout with password.
Calibration: complete instrument, output, and sensor calibration. Calibration can be NIST traceable.
Watchdog Timer: with a power supply monitor, to prevent unexpected instrument lockup.
Built-In Diagnostics: several self tests performed automatically and can be initiated at any time.
Specifications
Cond/Resis Ranges 0.002 S/cm to 600S/cm (using 0.01 and 0.1 cm-1 sensors)
1667 -cm to 500.0 M -cm
TDS: (ppk/ppm/ppb) covers equivalent conductivity range
Concentrations Ranges Not Available
Temperature -40 to 200C (-40 to 392F)
Resolution 0.001 S/cm, 0.001 M -cm, 0.010C
Sensor Inputs From Thornton conductivity sensors with Pt1000 RTD, via appropriate patch cord(s) or integral sensor cable (depends on sensor used).
Temperature Compensation Automatic, referenced to 25C for resistivity, conductivity, and TDS. Field selectable for Standard, Linear, Light84, Cation and None (Non temperature compensated measurement)
Outputs: Setpoints/Alarms; Relays Two controlled setpoints. Any relay can be activated by either setpoint: Two electro-mech SPDT, rated at 5 amp, 250 VAC or 30 VDC resistive load;Relays have individually adjustable delay and hysteresis (deadband).
Analog output SignalsTwo powered 4-20 mA outputs (recalibratable to 0-20 mA), 500 ohm load maximum, freely scalable to any parameter and range; isolated from input and from ground. Not for use with externally powered circuits.
Serial output RS232, maximum distance of 50 feet (15 m); RS422, maximum distance of 4000 feet (1220 m); field selectable up to 19.2 kbaud.
Performance
Accuracy 0.5% of reading or 0.5 ohm, whichever is greater; 0.25C
Repeatability 0.1% of reading, 0.13C
Update Rate All measurements and outputs, once per second
Ratings/approvals Meets CSA/NRTL and CE requirements, UL listed
Analog output accuracy 0.05 mA within 15-30C ambient
Environmental
Storage temperature -40 to 70C (-40 to 158F)
Operating temperature -10 to 55C (14 to 131F)
Humidity 0 to 95% RH, non-condensing
Enclosure
Display 16 character backlit LCD (4.8 x 9.6 mm)
Keypad 11 flush, tactile feedback keys
Power
Line 90-130 VAC or 180-250 VAC, 50-60 Hz, 12W maximum; or 12-30 VDC,300 mA steady state, 600 mA inrush. DC power must be isolated from earth ground and from other instruments
Memory retention On power loss, all programmed values are retained in non-volatile memory without batteries
Additional Prep Fees May Apply