The Agilent 1100 Series G1364C Automatic Fraction Collector has three operating modes that consist of: Discrete Fractions, which is the default mode for all vessels, where the flow is diverted to waste, while moving from one vessel position to the next vessel position; Continuous Low is available only when using the deep well plates; and Needle Into Location where the needle pushes into the vessel as deep as specified, for the use with capped vials and test tubes and well plates with closing mats.
Manufacturer Specification
Trigger Modes
Time slices, Peak (threshold, up- /downslope), Timetable (combination of time intervals and peak) and Manual trigger (supported only with G1323BControl Module)
Operating Modes
Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel position Continuous flow: optional, available only when using well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
Test Tube/Plate Sizes
Minimum 48 mm to 100 mm maximum
Maximum Tube Volume
ca. 45 ml
Maximum Flow Rate
100 ml / min (depending on viscosity and generated back pressure, max. 6 bar at the diverter valve)
Delay Volumes [l]
Fraction collector inlet to diverter valve: -500 (typical, depends on length of the tubing)
Diverter valve: ~15
Diverter valve to needle: ~110
Needle: -~5
Delay Calibration Sensor
Single wavelength absorbance detector working at 654 nm, consisting of a LED and a photo diode
Diverter Valve
3/2 Diverter valve with low internal volume (15 l), switching time < 100 ms, maximum operating pressure 6 bar
Maximum Capacity
3 fraction collectors in parallel plus one recovery fraction collector connected via 12-Position, 13-Port Selector valve
Interfaces
Controller-area network (CAN). optional; LAN or external contacts interface RS232C, APG-remote (for remote start / stop signals to / from other modules) Interface to G1330A Thermostat CAN-DC-out for operation of Agilent approved external devices like valves
Safety Features
Leak detection and safe leak handling, error detection and display, exhaust fan for fume extraction of hazardous vapors
Additional Prep Fees May Apply