Browse our other Testing and Analyzing Equipment
Get outstanding sensitivity for routine ion chromatography applications with the dual-lamp Thermo Scientific Dionex ICS-Series VWD Variable Wavelength Detector. Work free from worry of metal contamination with available all Thermo Scientific PEEK flow cells.Detector TypeVariable wavelength detector, single channelData Collection RateUp to 10HzAnalog OutputsTwo: absorbance, 20-bit resolution, 0-1V and 0-10VDrift<0.1mAU/hrNoise Level<2.5 AU at 254nmFlow Cell Volume11LWavelength Accuracy1nmWavelength Range190to900nmLampSimultaneous Tungsten and DeuteriumFlow CellPEEK, 11L volume, 10mm path length (analytical); 2.5L, 7mm path length (semimicro)Linearityup to 2.5AUPC ConnectionUSBWavelength Bandwidth6nm at 254nmHeight (Metric)15.2cmHeight (English)6.0 in.Hertz47/63HzVoltage85 to 265VACWidth (English)17.5 in.Width (Metric)44.4cmItem DescriptionSingle wavelengthElectrical Requirements85 to 265VAC 47/63HzEconomical High Performance for Routine ChromatographyExcellent sensitivity with low baseline drift and high signal-to-noise ratioLow noise and high intensity over the full spectral range with deuterium and tungsten lampsThermal stability with low backpressure from the integrated flow cell heat exchangerEasy verification of wavelength accuracy with built-in holmium oxide filterExcellent resolution that yields high linearityLamp lifetime monitor to prevent downtimeSimple installation and data collection via USBFlexible ApplicationsData collection rates up to 100HzBroad wavelength range of 190900nmPowerful, user-friendly control and data collection through Thermo Scientific Dionex Chromeleon Chromatography Data System softwareAdditional Prep Fees May Apply