Browse our other Testing and Analyzing Equipment
The LTR-25/140 temperature reference dry well is specifically designed to fully automate sensor calibration when used with Kaye validation software and traceable temperature standards.Kaye dry wells employ unique inserts which minimize cooling of the thermocouple tips due to heat conduction providing an uncertainty of merely 0.1C. Without proper inserts, transfer uncertainty in excess of 0.5C can occur with 22 gage, type T thermocouples.Manufacture SpecificationsTemperature Range-25C to 140CAmbient Operating Range5C to 50CSet-Pointing Accuracy0.2CTemperature Stability0.02CTransfer Calibration Accuracy*IRTD Standard to Thermocouples25C to 80C: 0.1C80C to 130C: 0.15C130C to 140C: 0.18CTypical Heat Up Time25C to 80C: 6 minutes25C to 140C: 14 minutesTypical Cool Down Time10 minutes: 25C to 20CDisplayLED w/0.01C resolutionComputer InterfaceRS232Additional Prep Fees May Apply