Thermo Scientific Velos Pro Mass Spectrometer System w/ Vacuum Pumps E2M30
***Unit was tested as far as we could. Unit is preowned with minimal wear from normal use. System will need to be validated/calibrated to ensure accurate results. Inquire if validation/calibration is available for an additional cost. Buyer can expect everything pictured with 30-day warranty. See photos for details.
(Please be advised that any warranty/guarantee/return policy provided from the seller is between the buyer and the seller only. Bid on Equipment is not and cannot be held responsible in any way. Please reference our Terms and Conditions for further information.)
Includes:
Velos Pro
Edwards E2M30 Vacuum Pumps (Qty: 2)
Ion Max API source
The Velos Pro mass spectrometer is an optimized dual-cell linear ion trap with high-pressure and low-pressure cells that blurs the lines between traditional ion trap and triple quadrupole performance.
The first high-pressure cell enables maximum ion trapping efficiency, higher ion isolation and fragmentation efficiency. The second low-pressure cell, enables better resolution and higher scan rates. All of which, results in higher sensitivity for full scan MS and MSn analysis.
If you are in the market for an extremely powerful and flexible mass spectrometer, look no further than the Velos Pro dual-cell linear ion trap.
Applications include:
proteomics, metabolite identification, forensic analysis, clinical research and teaching.
A single quadrupole offers no more than Selected Ion Monitoring (SIM). A triple quadrupole adds Selected Reaction Monitoring (SRM) where a selected ion can be fragmented into a product ion. A linear ion trap goes further by enabling Consecutive Reaction Monitoring (CRM) where a selected ion is fragmented into a product ion and the resulting product ion further fragmented in additional consecutive steps (MSn). Each additional fragmentation step makes compound identification more certain and facilitates structural characterization.
Hardware features
Thermo Scientific Ion Max API source
With ESI, HESI, APCI and APPI options
Sweep gas reduces chemical noise
60 interchangeable ion probe orientation
Removable metal ion transfer tube provides vent-free maintenance
Transfer ion optics
S-lens technology
Generation II Ion optics with novel neutral blocking technology for improved robustness
High stability and ion transmission efficiency
Dual-pressure 2D linear quadrupole ion trap mass spectrometer
Dual-pressure for optimized performance
Isolation waveforms during injection
Balanced rf field
Automatic system calibration
High-efficiency radial ion ejection
Vacuum system
Differentially-pumped vacuum system to 105 Torr
Split-flow turbomolecular pump controlling vacuum in three regions
Dual rotary vacuum pump configuration
High-vacuum aluminum analyzer chamber
Detection system
Proprietary dual conversion dynode detector
Two off-axis discrete dynode electron multipliers with extended dynamic range
24-bit electrometer for high level linearity
Digital electronic noise discrimination
Integrated liquid delivery
Fully-automated data system with valve control enables user to divert the solvent front, gradient end point, and any other portion of the HPLC run to waste
Syringe Pump allows automated infusion under data system control
Data system features
Hardware
High performance PC
27 inch LED Monitor with 1920 x 1080 resolution
Operating system
Microsoft Windows 10 x 64 enterprise IoT LTSB 2016 Thermo Scientific software
LTQ Instrument Control Software
Thermo Scientific Xcalibur software for method setup, data acquisition, data processing, and reporting
Thermo Scientific FreeStyle software to view and analyze data
Scan functions
Predictive Automatic Gain Control (AGC) delivers up to 10 Hz data dependent MS/MS acquisition
Full-scan mass spectra for sensitive analyses and rapid screening of unknown compounds
Full-scan product ion spectra at sensitivities higher than any ion trap mass spectrometer
Selected Reaction Monitoring (SRM) for traditional LC/ MS/MS quantitative analytical experiments
MSn for multi-stage MS experiments to probe the structure of ions
ZoomScan, a high-resolution, full-range scan to resolve isotopic envelopes often used for charge state
determination
Ultra ZoomScan for ultimate resolution
Rapid Scan, the fastest scan mode for UPLC analytical data collection
TurboScan an ultra-fast scan to improve signal-to-noise and sampling rate
Unique Automatic Gain Control (AGC) ensures that the ion trap is always filled with the optimum number of
ions for any scan type
Dynamic Exclusion allows acquisition of MS/MS and MSn spectra from lower intensity ion species
WideBand Activation generates more structurally informative spectras
Normalized Collision Energy provide reproducible data from instrument to instrument
System specifications
MS/MS sensitivity
Heated electrospray ionisation (HESI) 2 L of a 50 fg/L solution of reserpine (100 femtograms total)
injected at a flow of 400 L/min will produce a minimum signal-to-noise ratio of 100:1, for the transition of the
isolated protonated molecular ion at m/z 609 to the largest two product ions, m/z 397 and m/z 448, when
the mass spectrometer is operated at unit resolution in the full-scan MS/MS mode, scanning the product ion spectrum from m/z 165650.
Atmospheric pressure chemical ionization (APCI)
2 L of a 50 fg/L solution of reserpine (100 femtograms total) injected at a flow of 400 L/min will produce a minimum signal-to-noise ratio of 100:1, for the transition of the isolated protonated molecular ion at m/z 609 to the largest two product ions, 397 and 448, when the mass spectrometer is operated at unit resolution in the fullscan MS/MS mode, scanning the product ion spectrum from m/z 165650.
MSn sensitivity
Heated electrospray Ionization (HESI) A loop injection of 2 L of a 50 fg/L solution of reserpine (100 femtograms total sample) at a flow of 400 L/min will produce a minimum signal-to-noise ratio of 100:1, for the transition of the unit isolated protonated molecular ion at m/z 609 to the product ion at m/z 397 which is further fragmented to the product ion at m/z 365, when the mass spectrometer is operated at unit resolution in the full-scan MS/MS mode, scanning the product ion spectrum from m/z 165650.
Installation requirements
Power
One 230 Vac 10.0%, 15 Amps, 50/60 Hz, single phase, with earth ground dedicated to the instrument
120 or 230 Vac single phase, with earth ground for the data system
Gas
One high-purity (99% pure, flow rate 15 L/min) nitrogen gas supply for the API source
One ultra-high-purity helium gas supply (99.998% pure) with less than 1 ppm each of water, oxygen, and total hydrocarbons for the mass analyzer
Environment
System averages 2300 W (8000 BTU/h) output when considering air conditioning needs
Operating environment must be 1527 C (5980 F) and relative humidity must be 4080% with no condensation
Optimum operating temperature is 1821 C (6570 F)
Dimensions/weight
MS: 56 cm 79 cm 59 cm (h w d)
MS: ~120 kg
Two roughing pumps: 38.6 kg each
Performance specifications
Mass range
m/z 15200
m/z 502000
m/z 2004000
Linear dynamic range
Under typical experimental conditions, at least six (6) orders of magnitude
Polarity switching
100 msec between positive and negative
Scan power
MSn
or n = 1 through 10
Contact closure
Start In/Out
Start Out is programmable
Analog inputs
One (1) analog Input (01 V)
One (1) analog Input (010 V)
Additional Prep Fees May Apply