Thermo Scientific TriPlus 300 Headspace model: Master SHS/Master VH
The Thermo Scientific TriPlus 300 Headspace autosampler offers a comprehensive, flexible solution compatible with standard gas chromatographs (GC) and GC/MS systems, including those using packed and capillary columns, accelerating your organic volatiles determination workflow while minimizing down-time. The autosampler enables users to analyze a larger number of samples more quickly and achieve unparalleled productivity with enhanced sample tray and incubation oven capacity. The superior sample integrity and traceability, sample path inertness, seamless integration into multiple chromatography data systems, and immediate local user interface further expand the reliability, flexibility and applicability of the autosampler.
The ability to fully comply with existing methods in pharmaceutical, forensic and environmental laboratories makes this autosampler the product of choice for automated and dependable organic volatiles determinations. The wide range of settable temperatures for the entire sample path enables the TriPlus 300 HS autosampler to be delicate with biological samples and labile analytes and to be suitable for the fastest growing high-temperature headspace applications. Flexibility is maximized when the TriPlus 300 Headspace autosampler is used with Thermo Scientific TRACE 1300 Series GC instruments. The modularity of the latter lets the user move the autosampler from one GC to another in minutes, without modifying the GC pneumatics. This can be used to start a new application on a different GC, obtain confirmation on a GC-MS system, or just quickly satisfy incremental workloads. Priority samples can be placed at any sample tray position to rapidly answer urgent analysis needs.
Features
Injection Modes standards: Optimized to maximize vial overlapping to ensure the highest throughput
MHE (Multiple Headspace Extraction): Single and multiple puncture, with up to 100 extractions from a single vial
MHI (Multiple Headspace Injection): Up to 100 extractions for vial; GC run is started after the last injection
MDO (Method Developing Optimization): To optimize a headspace method, automatically incrementing one of the following parameters: oven temperature, thermostatting time, vial shaking level, number of MHI injections Other Injection Features
Automatic vial leak check; does not require calibrations and does not extend analysis time
Method actions: Skip, pause, continue, ignore, fake injection commands available for leak test, missing vial, wrong vial size and GC not ready
Automatic hardware diagnostic page, always showing system status logs and alarms for every vial.
Sound buzzer available to alert operator of sequence events or system error status.
Gas and Energy Saving becomes active after 10 minutes of complete inactivity when selected; automatic complete deactivation or user-selectable reduction of settings (temperature, gases and
touch screen intensity) to reduce the environmental impact.
Clock Events function allows the scheduling of unattended events (load a method, start a sequence, gas and energy saving) once, daily, weekly, or at a selectable date and time.
Automatic conditioning capabilities: User-settable time for conditioning heated zones, in order to obtain a higher temperature stability before a vial is loaded. Pneumatics
Equipped with two separate inlets on the rear for carrier gas and auxiliary gas
Carrier gas can be regulated by three different systems: Gas chromatograph, TriPlus 300 HS autosampler, or both of them.
Specifications
Compatible carrier gas: Hydrogen, helium, nitrogen
Auxiliary gas User-selectable: helium and nitrogen
Auxiliary gas electronic flow control Flow rate: 10 to 300 mL/min
Precision: +/- 0.1 mL/min
Safety control: On auxiliary gas presence
Maximum inlet pressure: 9 bar
Vial pressure: Up to 4 bar
Pressure settingsbar, psi, kPa
Vial vent Residual headspace vial pressure optionally eliminated via a vent port which can be connected to a trap or a fume hood if needed.
Carrier gas electronic control option Pressure range: 0120 psi
Carrier gas flow rate: max 1000 mL/min (He)
Compatible gas types: He, N2, H2, Ar, Ar+5%CH4
Compatible columns: Packed columns, wide-bore and capillary columns
Operation modes: TriPlus 300 HS control (direct control),
TriPlus 300 HS +GC mode (additional flow)
Pressure/flow programming: Constant or programmed flow, constant or programmed pressure.
Flow programming: 24 ramps
Pressure programming: 24 ramps
Pressure setpoint increment: 0.01 psi
Flow setpoint increment: 0.1 mL/min
Atmospheric pressure and
temperature compensation: Yes
Synchronization with GC oven: Up to 25 oven ramps and 26 isotherms
Oven unit/ Power Supply: 100120/ 200240 V~(10% max), 50/60 Hz, 1300 VA
Autosampler: 85264 V (autoranging)
Operating temperature range: 0 C to 40 C
Relative humidity range: 5 to 95% (without condensation)
Operating altitude: Up to 3500 m above sea level
Typical area repeatability: <1% RSD
Sample handling system: Three-axis robot autosampler
Vial size: Without the use of any vial-height adapter, TriPlus 300 Headspace autosampler is compatible with 10, 20, and 22 mL headspace vials with crimped or screwed magnetic caps with flat or rounded bottom style
Vial capacity: 120 positions standard. Optional cooled tray for 120 positions, which requires an external circulator bath;
optional tray for 60 vials positions.
Oven capacity: 18-place electrically driven carousel
Sample overlapping: Yes, with up to 18 vials; timing automatically optimized to keep the equilibration time constant, maximizing throughput. Sample overlapping works also on non-consecutive vial positions of the sample tray.
Oven Settings: Temperature: OFF or from 30 C to 300 C, settable in 1 C increments, and with 0.1 C read out
Equilibration time: 0 to 999 min, increment 0.01 min
Vial shaking: None, slow, medium, high
Switching valve: 6 ports, electrically actuated; Temperature: OFF or from 30 C to 300 C, settable in 1 C increments, and with 0.1 C read out
Transfer line temperature: OFF or from 30 C to 300 C, settable in 1 C increments and with 0.1 C read out
Timing Pressure equilibration time: User settable up to 5 minutes
Vial equilibration time: Up to 999 minutes
Loop fill time: Up to 5 minutes
Loop equilibration time: Up to 999 minutes
Injection time: Up to 999 minutes
GC time: Up to 999 minutes
Priority vials and sequences: Yes, indefinite numbe
Vial pressurization modes: Standard, Pressure, Flow, Volume. All are fully controlled by the auxiliary gas electronic flow control on the unit
Loop fill modes: Standard, Pressure, Custom. All are fully controlled by the auxiliary gas electronic flow control on the unit
Purging: Automated purging of vent and sample paths, fully controlled by the auxiliary gas electronic flow control on the unit. User-selectable settings for time and purge flow
Loop sample path: Stainless steel coated with inert material (SilcoNert 2000)
Standard loop: 1 mL; optional loops: 0.5, 3 and 5 mL
Sample pathway and needle: All sample flow path is stainless steel coated with chemically inert material (SilcoNert 2000)
Transfer line: Stainless steel internally coated with inert material (SilcoNert 2000) Inner diameter: 1.00 mm Length: 80 cm standard; optional extended 1.05 meter transfer line
Additional Prep Fees May Apply